| Model Number |
URF-P7 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the uretero reno fiberscope exhibited foreign objects inside the objective lens.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Based on the results of the investigation, olympus confirmed foreign material adhering to the objective lens of the distal end section.It was not possible to identify the foreign matter.There was no physical damage at the place where the foreign material remained.The reprocessing information was unavailable, therefore; the legal manufacturer was unable to confirm if there were deviations from the reprocessing steps as presented in the instructions for use.A definitive root cause could not be determined.The event can be detected and prevented in accordance with the following ifu.The instruction manual operation manual ¿urf-p7, chapter 3 preparation and inspection¿ describes the methods for detection.The instruction manual reprocessing manual ¿urf-p7, chapter 5 reprocessing the endoscope¿ describes the methods for prevention.Olympus will continue to monitor field performance for this device.
|
| |
|