| Model Number |
LTF-190-10-3D |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the endoeye 3d flex videoscope did not display an image during preparation for usage in an unspecified procedure.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the investigation results, it is presumed that water/moisture invaded the device and image sensor unit / printed circuit board inside connector was broken by the water, causing the suggested event.The event can be detected by following the instructions for use (ifu): ltf-190-10-3d operation manual chapter 3 preparation and inspection:3.6 inspection of the endoscopic system: inspection of the endoscopic image.Olympus will continue to monitor field performance for this device.
|
| |
|