• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET DEPTH GAUGE 70MM; KNEE INSTRUMENT: DEPTH GUAGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET DEPTH GAUGE 70MM; KNEE INSTRUMENT: DEPTH GUAGE Back to Search Results
Catalog Number 227460000
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that the 16 mm standard distal reamer splines were damaged when removing the trial neck from the reamer.The 51/28 mobile bearing head trial was dropped on the floor and lost by hospital staff.Also, the depth gauge was accidently damaged and bent.None of these issues delayed surgery.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot
=
> the device lot number is unknown, therefore a¿device history review could not be performed.¿ if the lot/serial number becomes available, the record will be re-assessed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 lot, d9, h4.Corrected: d3, d4 primary udi number.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) investigation summary it was reported that the 16 mm standard distal reamer splines were damaged when removing the trial neck from the reamer.The 51/28 mobile bearing head trial was dropped on the floor and lost by hospital staff.Also, the depth gauge was accidently damaged and bent.None of these issues delayed surgery.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device found that the quickset depth gauge 70mm was severely deformed from the tip.Potential cause can be attributed to unintended use error by heavy usage and of excessive force in a prying motion and overload of the material.Properly handling and attention to the approved use of the device diminishes the risk of failure.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the quickset depth gauge 70mm would have contributed to the complained issue.¿based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET DEPTH GAUGE 70MM
Common Device Name
KNEE INSTRUMENT: DEPTH GUAGE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20391333
Report Number1818910-2024-21018
Device Sequence Number15102385
Product Code HTJ
UDI-Device Identifier10603295109358
UDI-Public(01)10603295109358(10)SO2044848
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number227460000
Device Lot NumberSO2044848
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/01/2024
Supplement Date Received by Manufacturer10/29/2024
01/08/2025
02/06/2025
Initial Report FDA Received Date10/07/2024
Supplement Report FDA Received Date10/30/2024
01/10/2025
02/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-