| Model Number |
8881850310 |
| Medical Device Problem Codes |
Break (1069); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported that the needle hub broke which resulted in leakage at the luer connection.The syringe could not be used due to a product leak.A slight cracking noise was heard when the cannula was placed on the syringe.The syringe leaked.There is a crack in the needle tip between the needle and the screw thread.The syringe broke between the needle and the start of the syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Samples and pictures were received, and the reported issue was observed.However, a definitive root cause could not be determined at this time.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|