|
This report is being supplemented to provide additional information based on the approved final investigation.New information added: d8, h4: the subject device was manufactured in apr.2024, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the subject device was not returned to aomori olympus.Therefore, the condition of the device could not be confirmed.In-house inventory guide wire distal end found that its strength did not meet specifications, and the breakage pattern showed tearing, similar to the product in the reported event.Inspection of the fracture surface of the low-strength guide wire distal end revealed layered resin solidification and the presence of welds.Brittle fractures originated at the welds, making the guide wire distal end more susceptible to tearing.This suggests a molding defect in the guide wire distal end, indicating that it was not molded under optimal processing conditions.The cause of this event was that the guide wire distal end did not meet strength specifications due to suboptimal molding conditions during manufacturing.A definitive root cause of the reported issue could not be determined.The instruction manual contains a warning to the user regarding this event.When using the guide wire, insert the instrument with its distal tip in parallel with the guide wire while holding the distal tip as shown in figure 4.Be careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire as shown in figure 5.This may damage the distal tip.Olympus will continue to monitor field performance for this device.
|