• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH URETEROSCOPE 7.3/10.4 FR. X 430 MM, 5°, ANGLED OCULAR, 6.4 FR. CHANNEL; RIGID URETEROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH URETEROSCOPE 7.3/10.4 FR. X 430 MM, 5°, ANGLED OCULAR, 6.4 FR. CHANNEL; RIGID URETEROSCOPE Back to Search Results
Model Number WA2UR14A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was observed that during the device inspection, the rigid ureteroscope's handle was broken.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additional information: h3, h4, h6, h11 a review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event is most likely due to the effect of a large mechanical force due to a shock, fall or collision with other equipment.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETEROSCOPE 7.3/10.4 FR. X 430 MM, 5°, ANGLED OCULAR, 6.4 FR. CHANNEL
Common Device Name
RIGID URETEROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20396204
Report Number9610773-2024-02468
Device Sequence Number12368579
Product Code FGB
UDI-Device Identifier04042761086363
UDI-Public04042761086363
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K200369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberWA2UR14A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2024
Supplement Date Received by Manufacturer01/17/2025
Initial Report FDA Received Date10/08/2024
Supplement Report FDA Received Date01/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-