• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PRONOVA BLU 36IN 4-0 D/A RB-1; SUTURE, NON-ABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PRONOVA BLU 36IN 4-0 D/A RB-1; SUTURE, NON-ABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Catalog Number PN3557H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 10/8/2024.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The lot/batch was not provided; therefore, a manufacturing record evaluation could not be performed.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: please provide lot number of the devices involved in the reported event.Please confirm the quantity of devices involved in the reported event and the quantity of devices available to be returned.Please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq).Please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent a hepato-biliary-pancreatic surgery on (b)(6) 2024 and suture was used.During a hepato-biliary-pancreatic surgery, the needles had been detached from the sutures several times.It was not a control release needle.There were no adverse consequences to the patient.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 10/28/2024.Additional information was requested, the following was obtained: - please provide lot number of the devices involved in the reported event.=>unk.- please confirm the quantity of devices involved in the reported event and the quantity of devices available to be returned.- please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq).=>(b)(6).- please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.=>we regularly contact with sales rep about the device returning.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).D4: full udi currently unavailable since the exact lot of the suspected device cannot be determined.Additional information: d4 ¿ the actual device batch number associated with this event is not known.The following are the possible batch numbers: slbhxm, tgbjxs, ubbhbq, tmbepq, tmbdum, tmbcbu, slbcup, ubbhbq.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.1.Are all of the lot numbers possible lots for each impacted product? 2.What lot numbers are associated with each impacted product? 3.Please clarify the quantity that is being returned for each impacted product.This report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 2/20/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Additional information.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: 1.Are all of the lot numbers possible lots for each impacted product? yes 2.What lot numbers are associated with each impacted product? unk 3.Please clarify the quantity that is being returned for each impacted product.Slbhxm x6, tgbjxs x2, ubbhbq x3 and tmbepq x1.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with twenty-six representative unopened samples was received for analysis.Product code pn3557h, lot slbhxm.Upon initial inspection of the samples, no external damages were observed on the external packets.As per the sampling plan, a visual inspection was performed on thirteen samples.The packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.H3 investigational summary: the product was returned to ethicon for evaluation.Three representative unopened samples were received for analysis.Product code pn3557h, lot tmbcbu.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with nineteen representative unopened samples was received for analysis.Product code pn3557h, lot ubbhbq.Upon initial inspection of the samples, no external damages were observed on the external packets.As per the sampling plan, a visual inspection was performed on thirteen samples.The packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.H3 investigational summary: the product was returned to ethicon for evaluation.Four representative unopened samples were received for analysis.Product code pn3557h, lot tmbepq.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.H3 investigational summary: the product was returned to ethicon for evaluation.Eight representative unopened samples were received for analysis.Product code pn3557h, lot tmbdum.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.H3 investigational summary: the product was returned to ethicon for evaluation.One representative unopened sample was received for analysis.Product code pn3557h, lot tgbjxs.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned sample, the packet was opened, and no defects were detected.The swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.H3 investigational summary: the product was returned to ethicon for evaluation.One representative unopened sample was received for analysis.Product code pn3557h, lot slbcup.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned sample, the packet was opened, and no defects were detected.The swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional information: d4, h4 - the actual device lot number associated with this event is not known.The possible lot numbers are reported as follows: batch ubbhbq, batch slbcup, batch tmbcbu, batch tmbdum, batch tmbepq, batch ubbhbq, batch tgbjxs, batch slbhxm.A manufacturing record evaluation was performed for the finished device batch slbhxm, pn3557h and no non-conformances were identified.Mfg.Date - 10/27/2022, exp.Date - 09/30/2027.A manufacturing record evaluation was performed for the finished device batch tgbjxs, pn3557h and no non-conformances were identified.Mfg.Date - 06/27/2023, exp.Date - 05/31/2028.A manufacturing record evaluation was performed for the finished device batch ubbhbq, pn3557h and no non-conformances were identified.Mfg.Date - 02/21/2024, exp.Date - 01/31/2029.A manufacturing record evaluation was performed for the finished device batch tmbepq, pn3557h and no non-conformances were identified.Mfg.Date - 11/20/2023, exp.Date - 10/31/2028.A manufacturing record evaluation was performed for the finished device batch tmbdum, pn3557h and no non-conformances were identified.Mfg.Date - 11/16/2023, exp.Date - 10/31/2028.A manufacturing record evaluation was performed for the finished device batch tmbcbu, pn3557h and no non-conformances were identified.Mfg.Date - 11/08/2023, exp.Date - 10/31/2028.A manufacturing record evaluation was performed for the finished device batch slbcup, pn3557h and no non-conformances were identified.Mfg.Date - 10/13/2022, exp.Date - 09/30/2027.A manufacturing record evaluation was performed for the finished device batch ubbhbq, pn3557h and no non-conformances were identified.Mfg.Date - 02/21/2024, exp.Date - 01/31/2029.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRONOVA BLU 36IN 4-0 D/A RB-1
Common Device Name
SUTURE, NON-ABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20396497
Report Number2210968-2024-10598
Device Sequence Number6517495
Product Code GAT
UDI-Device Identifier10705031246164
UDI-Public10705031246164
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K001625
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPN3557H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2024
Supplement Date Received by Manufacturer10/16/2024
12/24/2024
01/24/2025
Initial Report FDA Received Date10/08/2024
Supplement Report FDA Received Date10/28/2024
01/08/2025
02/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-