| Catalog Number |
4826162 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.Donor unit id #: (b)(4) there was not a transfusion recipient or patient involved at the time of the residual wbc testing; therefore, no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly the result of a wbc saturation of the lrs chamber during the procedure which fell just under the detection threshold for flagging the platelet product for wbc verification.Based on the available information, it is likely that the higher-than-expected wbc content in the platelet product may have been donor-related.Suggestions you may consider monitoring this donor's next few platelet donations for wbcs to determine if there is any donor specific leukoreduction failure trend.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Investigation is in process; a follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.Donor unit id #: (b)(6).There was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly the result of a wbc saturation of the lrs chamber during the procedure which fell just under the detection threshold for flagging the platelet product for wbc verification.Based on the available information, it is likely that the higher-than-expected wbc content in the platelet product may have been donor-related.Suggestions you may consider monitoring this donor's next few platelet donations for wbcs to determine if there is any donor specific leukoreduction failure trend.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: a root cause assessment was performed for this complaint.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was possibly the result of a wbc saturation of the lrs chamber during the procedure which fell just under the detection threshold for flagging the platelet product for wbc verification.Based on the available information, it is likely that the higher-than-expected wbc content in the platelet product may have been donor-related.
|
| |
|