| Model Number |
1012272-12 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.The investigation determined that the reported complaints appear to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the heavily calcified anatomy resulting in the reported difficulty advancing the device and subsequently the balloon became damaged and ruptured during inflation.Based on the reported information, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 90% stenosed de novo lesion in the right coronary artery (rca) with heavy calcification and mild tortuosity.The 2.5x12mm trek balloon dilatation catheter (bdc) was advanced for pre-dilatation and with resistance was noted due to the anatomy.The balloon was inflated to 14 atmospheres (atm); however, the balloon ruptured during the first inflation.The bdc was removed from the patient and another trek balloon was used to continue the procedure.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|