|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3543751 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/31/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Date of explantation: (b)(6) 2024.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation primary with mentor memorygel breast implants 375cc and experienced bilateral capsular contracture baker grade iii post-operatively, which was confirmed during a physical exam.As a result, the patient is scheduled for removal on (b)(6) 2024.This report is for the left breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On oct 15, 2024, additional information was received indicating the patient underwent removal and replacement with mentor gel breast implants (serial numbers (b)(6)) on (b)(6) 2024.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On nov 8, 2024, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample determined that the sx mpp gel 375cc breast implant had an area of separated material on the anterior view, measuring approximately 3.5 cm x 2.0 cm.Microscopic examination was performed on the edges of the separated material, and parallel striations were found in an area of the tear, measuring approximately 0.1 cm.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.The cause of the rupture in the remaining area of the separated material could not be identified.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body's individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|