• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 REVERSE LINER 40+0 R; SHOULDER IMPLANT - HUMERAL CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 REVERSE LINER 40+0 R; SHOULDER IMPLANT - HUMERAL CUP Back to Search Results
Catalog Number 550040100
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 09/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient claims that they dislocated their shoulder by sweeping and had several previous instances of what felt like dislocation prior to the complete dislocation.Doi: (b)(6) 2024 dor: (b)(6) 2024 affected side: left shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : it was reported that the patient claims that they dislocated their shoulder by sweeping and had several previous instances of what felt like dislocation prior to the complete dislocation.Unknown surgical delay.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found nothing indicative of a device nonconformance.The device exhibits scratches related to the extraction process.The allegation of implant dislocation can not be confirmed as no patient x-rays were provided to assess the condition.A search of the j&j medtech orthopaedics nonconformance (nc) quality system found no nc¿s associated with this product 550040100/lot mi125764 combination.A functional test was not conducted due to not being applicable to the complaint condition a dimensional inspection was not conducted due to not being applicable to the complaint condition the overall complaint was not confirmed as the observed condition of the reverse liner 40+0 r would have not contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A search of the j&j medtech orthopaedics nonconformance (nc) quality system found no nc¿s associated with this product 550040100/lot mi125764 combination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE LINER 40+0 R
Common Device Name
SHOULDER IMPLANT - HUMERAL CUP
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20398267
Report Number1818910-2024-21070
Device Sequence Number20453001
Product Code PHX
UDI-Device Identifier10603295546542
UDI-Public(01)10603295546542(17)270331(10)MI125764
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K212737
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number550040100
Device Lot NumberMI125764
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/27/2024
Supplement Date Received by Manufacturer10/22/2024
11/25/2024
Initial Report FDA Received Date10/08/2024
Supplement Report FDA Received Date10/23/2024
11/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REVERSE LINER 40+0 R
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-