The device was not returned for evaluation.Lot history record (lhr) and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.The emboshield nav6 instruction for use (eifu), ¿do not attempt to move the filtration element after deployment and prior to the start of retrieval.¿ based on the reported information, there was no device issue reported with the emboshield nav6 device.Based on the reported information, the reported difficulties appear to be due to inadvertent retraction of the barewire during the procedure causing the filter to move down the carotid artery.There is no indication of a product quality issue with respect to manufacture, design, or labeling.The additional device referenced in b5 is filed under a separate medwatch report number.H6: medical device problem code 2017 - incorrect removal.
|