| Model Number |
CTF73 |
| Medical Device Problem Codes |
Break (1069); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event unit is anticipated to return to applied medical for evaluation.A follow-up report will be sent upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: unilateral adrenalectomy in prone position.Event description: complaint (b)(4) (1 of 2), complaint (b)(4) (2 of 2).The trocar broke in the middle part during the operation and was discovered on removal when half of the trocar got stuck in the port hole.It is unclear when during the operation the trocar broke.Two different article numbers are sent in by the customer.The customer will send in 2 art no (cts22 and ctf73) lot no 1525327 and lot no 1514216.Additional information received from an applied medical representative a email on 27sep24: the tm had spoken with the nurse in charge today earlier to confirm the following - both devices broke during the same procedure on (b)(6) 2024.Aware date was updated on 26sep24 for both complaints.No patient injury and broke piece was retrieved after surgery.Intervention: broke piece was retrieved after surgery.Patient status: no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of a damaged trocar.The wall thickness of the cannula shaft near the fracture was measured and did not meet specification.Based on the description of the event and condition of the returned unit, it is possible the cannula shaft bent as a result of the insertion force or device manipulation after insertion.It is also possible the uneven wall thickness made the cannula more susceptible to bend.Applied medical has performed a historical trend analysis and review of production records and no relevant documentation was identified.This event was reported based on the event description of a broken trocar.However, based on further evaluation of the returned unit, applied medical determined that this event is not reportable as a bent cannula shaft is unlikely to cause or contribute to death or serious injury.[correction] sections d1 and d4 (primary unique device identification number and additional unique device identification number) have been updated.
|
| |
|
Event or Problem Description
|
|
Procedure performed: unilateral adrenalectomy in prone position.Event description: complaint (b)(4) (1 of 2), mfr report number 2027111-2024-00863.Complaint (b)(4) (2 of 2), mfr report number 2027111-2024-00863.The trocar broke in the middle part during the operation and was discovered on removal when half of the trocar got stuck in the port hole.It is unclear when during the operation the trocar broke.Two different article numbers are sent in by the customer.The customer will send in 2 art no (cts22 and ctf73) lot no 1525327 and lot no 1514216.Additional information received from an applied medical representative via email on 27sep24: the applied medical representative had spoken with the nurse in charge today earlier to confirm the following - both devices broke during the same procedure on (b)(6) 2024.Aware date was updated on 26sep24 for both complaints.No patient injury and broke piece was retrieved after surgery.Additional information received from an applied medical representative via email on 01oct24: no patient was harmed and not a robotic surgery.Intervention: broke piece was retrieved after surgery.Patient status: no patient injury.
|
| |
|