• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION P500 SURFACE; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION P500 SURFACE; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P005787
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The baxter technician found the power cord needed to be replaced.It is necessary for the p500 to have an effective maintenance program.We recommend that you do annual preventive maintenance pm and testing for joint commission certification.Pm and testing not only meet joint commission requirements but can help make sure of a long, operative life for the p500.Pm will minimize downtime due to excessive wear.Examine the plug for damage.Make sure the plug is a onepiece molded plug assembly.If it is not, replace the plug cord assembly.Replace any plug cord assembly that shows any of these: discoloration of the plug molding around the plug blades, this could occur if the plug blades have overheated or arced.Any signs of cracking, this could occur if the plug has been bent and straightened to a point past its useful life.Loos fit of the plug blade the plug blade moves in the molding, this could occur if the molding has overheated or the blades have been bent and straightened to a point past their useful life.Replace the power cord, if damaged.A search of the baxter maintenance records showed baxter performed preventive maintenance on this bed in feb 9,2024.It is unknown if the facility performed any other preventive maintenance on this bed.The technician replaced the power cord to resolve the reported event.Based on this information, no further action is required.
 
Event or Problem Description
Baxter received a report from a baxter technician to report that power cord was exposing copper wires.The bed was located at the account and occurred before use.There was no patient/user injury, medical intervention, symptom, or adverse event reported.This report was filed in our complaint handling system as complaint pr # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P500 SURFACE
Common Device Name
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key20403552
Report Number1824206-2024-01430
Device Sequence Number3351805
Product Code FNL
UDI-Device Identifier00887761001022
UDI-Public(01)00887761001022(11)110106
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberP005787
Device Catalogue NumberP005787RENT02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2024
Initial Report FDA Received Date10/08/2024
Date Device Manufactured06/01/2011
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Patient SexUnknown
-
-