• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; NASAL TUBING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; NASAL TUBING Back to Search Results
Model Number BC190-05
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Section d4: serial number intentionally left blank as the information is not applicable section g4: no 510(k) number was included due to the subject device being a class 1 device, therefore, exempt from pma pathway to regulatory approval method: the complaint bc190 flexitrunk infant nasal tubing was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the photograph and description of events provided by the customer, previous investigations of similar complaints and our knowledge of the product.Results: visual inspection of the provided photograph revealed that the tubing of the bc190 flexitrunk infant nasal tubing was observed to be torn between the 8th and 9th filament from the midline connector, confirming the reported fault.Conclusion: without the return of the subject device, our investigation was unable to determine the exact cause of the observed damage.All bc190 bubble cpap system are visually inspected, and pressure tested before leaving the production line, those that fail are rejected.This suggests that the damage on the subject nasal tubing occurred after it was released for distribution.Our user instructions which accompany the bc190 bubble cpap system state: - "use caution when positioning or disconnecting to the infant interface.Avoid excessive pull forces, sharp objects, and tubing holders.Damage to the tubing may cause loss of pressure and require immediate replacement." a caution insert is included in the packaging to remind the user to take extra care when handling the bc190 bubble cpap system.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in china that the tubing of a bc190 flexitrunk infant nasal tubing was found damaged before patient use.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
NASAL TUBING
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key20403769
Report Number9611451-2024-00718
Device Sequence Number3379238
Product Code BZO
UDI-Device Identifier09420012413390
UDI-Public(01)09420012413390(10)2102892063(11)231121(17)261121
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBC190-05
Device Catalogue NumberBC190-05
Device Lot Number2102892063
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/11/2024
Initial Report FDA Received Date10/08/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-