Model Number 7122/60
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via product receipt that the patient had developed infection.The whole system including the implantable cardioverter defibrillator, right ventricular lead, right atrial lead, left ventricular lead, and left bundle lead, was explanted and replaced with a temporary pacemaker system.Patient condition was stable throughout.
 
Event or Problem Description
It was reported via product receipt that the patient had developed infection with no allegation of it being procedure or product related, except for a "poor pocket." the whole system including the implantable cardioverter defibrillator, right ventricular lead, right atrial lead, left ventricular lead, and left bundle lead, was explanted and replaced with a temporary pacemaker system.Patient condition was stable throughout.
 
Additional Manufacturer Narrative
A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The product was returned and visual inspection was normal.The cause of infection could not be traced to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key20404020
Report Number2017865-2024-66884
Device Sequence Number3379244
Product Code LWS
UDI-Device Identifier05414734502146
UDI-Public(01)05414734502146(10)A000157725(17)270430
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number7122/60
Device Lot NumberA000157725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2024
Supplement Date Received by Manufacturer10/10/2024
Initial Report FDA Received Date10/08/2024
Supplement Report FDA Received Date10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-