| Model Number |
AA10S |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Repair request initiated for device with the report of overheating.No patient impact reported.Repair request escalated to product event due to customer indication that this report is a complaint.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 product analysis: device returned; evaluation anticipated but not yet started.B3 date of event notification: 13-sep-2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation could not reproduce the reported malfunction as the temperature was within specification at 95°f.However, it was noted that inadequate preventative maintenance.This finding may have contributed to the user¿s experience.It was noted that bearing worn, the color ring is scored, the laser markings are faded, some internal tube's items are stuck on it.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|