• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752181
Medical Device Problem Codes Obstruction of Flow (2423); Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that just after ultra sound mode started, an occlusive sound kept ringing and aspiration could not be performed at all during cataract surgery.Despite connected an irrigation and aspiration tube and performed a priming again, the situation did not improved at all.The surgery was resumed and completed after replacing the pack with another one.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The used fluidics management system (fms) in the tray was visually inspected and no obvious defects were observed.Both irrigation and aspiration connectors were intact.The sample was tested using the console.The sample could be recognized by the console.The sample primed and tuned with the tip and infusion sleeve from the lab stock successfully.No system message code displayed on console screen and the service data could be retrieved.The sample functioned within specification.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.No contributing factors could be identified that could cause the customer¿s experience.After an investigation of this complaint, it has determined that no further actions will be pursued at this time for this event as the returned sample functioned per specifications.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20405809
Report Number1644019-2024-02324
Device Sequence Number6517852
Product Code LRO
UDI-Device Identifier00380657521814
UDI-Public(01)00380657521814(17)251031(11)231110(10)16A11V
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Catalogue Number8065752181
Device Lot Number16A11V
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/12/2024
Supplement Date Received by Manufacturer12/12/2024
Initial Report FDA Received Date10/09/2024
Supplement Report FDA Received Date01/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION VISION SYSTEM
Patient SexUnknown
-
-