Block a3b: the patient's gender is unknown.This information was not available from the facility.Block b6: patient information regarding relevant tests/laboratory data is unknown.This information was not available from the facility.Block h3: the stellarex device was returned for evaluation.Visual inspection found a pinhole on the distal side of the balloon.During functional testing, using a syringe filled with water, the balloon was pressurized and at 1 atm, a pinhole leak was confirmed at the distal side of the balloon.Block h6: per the ifu, balloon rupture is listed as a potential complication of the peripheral balloon.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|