| Model Number |
TJF-145 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the duodenovideoscope exhibited foreign material in the light guide lens.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.New information has been added to the following fields: d8, h3, h4, h6, h11.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, deformation of the nozzle was observed as a trace that physical stress was applied to the subject device.A definitive root cause cannot be determined.As a result of checking the instructions for use and labeling of the products, there were no abnormalities that could lead to the phenomenon.Olympus will continue to monitor field performance for this device.
|
| |
|