• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. CEMENTED FINNED TIB. TRA SZ 3F/3T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CEMENTED FINNED TIB. TRA SZ 3F/3T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-04-33
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/03/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitants: ((b)(6)), 200-04-33 - cemented finned tib.Tra sz 3f/3t; ((b)(6)), 200-02-35 - three peg patella 35mm; ((b)(6)), 200-01-03 - cemented cr femoral sz 3.The product associated with the reported event is within the scope of recall z-0019-2022.However, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
Legal case ¿ usa: it was reported that approximately 203 months after a left total knee replacement procedure, the patient underwent a revision procedure to address loosening.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.The serial number 0808109 is confirmed to be a part of recall number: z-0019-2022.
 
Additional Manufacturer Narrative
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.D4: corrected.D10: concomitant devices: (b)(6) 200-23-09 - cr tibial insert sz3, 9mm, (b)(6) 200-02-35 - three peg patella 35mm, (b)(6) 200-01-03 - cemented cr femoral sz 3, g4: corrected, h6: corrected component, and investigation clinical codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CEMENTED FINNED TIB. TRA SZ 3F/3T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key20408090
Report Number1038671-2024-03896
Device Sequence Number6518395
Product Code JWH
UDI-Device Identifier10885862039736
UDI-Public10885862039736
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K932776
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/04/2010
Device Catalogue Number200-04-33
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2024
Supplement Date Received by Manufacturer05/13/2025
Initial Report FDA Received Date10/09/2024
Supplement Report FDA Received Date05/13/2025
Date Device Manufactured10/28/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexMale
-
-