| Model Number |
FIBERWIRE #1, BLUE, 38" |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, it was reported by a sales representative via email that an ar-4068hl swiftstitch suture passer frayed the suture.This was discovered during a procedure.Additional information received on 9/19/2024: this was discovered during a hip arthroscopy labral repair procedure on 9/12/2024.The swiftstitch suture passer frayed an ar-7216 fiberwire, which broke.Everything was removed from inside the patient.The case was completed using products from another manufacturer.It was delayed about two minutes without additional anesthesia.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d6a, g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
|
| |
|