| Catalog Number |
5530-G-309-E |
| Medical Device Problem Codes |
Material Erosion (1214); Fracture (1260)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The insert has fractured necessitating a revision and replacement of the insert.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding crack/fracture and wear involving a triathlon insert was reported.The event was confirmed via provided photographs of the device.Method & results: device evaluation and results: the reported device was not returned however photographs were provided for review.The photographs show a recently explanted insert with wear on the posterior-medial edge.Two fractured fragments of the device are also visible.Clinician review: no medical records were received for review with a clinical consultant.Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: it was reported that the patient was revised due to fracture of the insert.The reported device was not returned however photographs were provided for review.The photographs show a recently explanted insert with slight wear on the posterior-medial edge.Two fractured fragments of the device are also visible.The exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|
Event or Problem Description
|
|
The insert has fractured necessitating a revision and replacement of the insert.
|
| |
|