• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 5.4 MM, 30°, AUTOCLAVABLE; WA50373B

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 5.4 MM, 30°, AUTOCLAVABLE; WA50373B Back to Search Results
Model Number WA50373B
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1.Full establishment name: (b)(6) hospital the evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported, the telescope, 5.4 mm, 30°, autoclavable had damage to the light source cord.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additional information: b5, d8, d10, h3, h4, h6, h11.The reportable malfunction was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the damage can be attributed to improper handling and application of excessive force, impact to the device like fall, shock or similar stress.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The issue occurred during reprocessing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 5.4 MM, 30°, AUTOCLAVABLE
Common Device Name
WA50373B
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20409220
Report Number9610773-2024-02488
Device Sequence Number21093410
Product Code GCJ
UDI-Device Identifier04042761029582
UDI-Public04042761029582
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K923982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA50373B
Device Lot Number501469
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Supplement Date Received by Manufacturer08/06/2025
Initial Report FDA Received Date10/09/2024
Supplement Report FDA Received Date08/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
WA03310A - LIGHT GUIDE CABLE
Patient SexUnknown
-
-