Model Number IS3-WW
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the trocar broke during attempt to implant the first stent from a trabecular microbypass stent system in the patient's right eye (od).Per report, the trocar fragment was removed from the eye intraoperatively, and a back-up device was used to complete the procedure.The report noted the event as "resolved" with no patient injury.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information: a6: race noted as chinese.The device was not available; therefore, product testing on the actual device could not be performed.The device history record review of the manufacturing lot was performed and there were no non-conformities found to be related to the reported event.A complaint history review was completed.The electronic complaint system was searched for the specified lot#.There were no similar issues reported for this lot #.A review of the device labeling including the risk analysis was completed.The reported issue (trocar fragmentation) is identified to be an anticipated hazard.The residual risk for each potential harm was found to be within the insignificant risk region.The ifu adequately provides instructions for stent implantation, precautions, and warnings for the proper use and handling of the device.Mfr# reference:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM
Common Device Name
AQUEOUS SHUNT
Manufacturer (Section D)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer (Section G)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer Contact
brittany massey
229 avenida fabricante
san clemente, CA 92672
949367960
MDR Report Key20409225
Report Number2032546-2024-00116
Device Sequence Number9044110
Product Code KYF
UDI-Device Identifier00853704002470
UDI-Public01008537040024701032125917260630213
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K220032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model NumberIS3-WW
Device Lot Number321259
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2024
Initial Report FDA Received Date10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient RaceAsian
-
-