• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 BD PREVUE II PERIPHERAL VASCULAR ACCESS SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 BD PREVUE II PERIPHERAL VASCULAR ACCESS SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770700
Medical Device Problem Codes Poor Quality Image (1408); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the image does not provide adequate visualization of the vessel in order to place catheter.Using this machine is leading to interruptions in patient care and patient safety concerns.No other information provided, at this time.
 
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and is pending evaluation.A supplemental will be submitted with evaluation results.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was returned to the service facility for evaluation.During evaluation, the reported issue of the image does not provide adequate visualization of the vessel in order to place catheter was unconfirmed.The prevue ii scanner is functioning properly within manufacturer specifications.The reported issue root cause is not required due to the reported issue is unconfirmed.
 
Event or Problem Description
It was reported that the image does not provide adequate visualization of the vessel in order to place catheter.Using this machine is leading to interruptions in patient care and patient safety concerns.No other information provided, at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PREVUE II PERIPHERAL VASCULAR ACCESS SYSTEM
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
shelly gilbert
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key20410019
Report Number3006260740-2024-05797
Device Sequence Number9104858
Product Code IYO
UDI-Device Identifier00801741214912
UDI-Public(01)00801741214912(11)240301
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K211193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9770700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2024
Supplement Date Received by Manufacturer11/13/2024
Initial Report FDA Received Date10/09/2024
Supplement Report FDA Received Date12/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-