Catalog Number SC60A
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 10/9/2024.D4 batch #529c84.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the sc60a device was received for analysis with no apparent damaged and with a gst60b reload loaded on the device.The reload was received partially fired 1/2.The device was tested for functionality in the articulated position with the returned reload by resetting and reloading it into the device.The device achieved its complete stroke firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.The partially fired is consistent with an incomplete or interrupted cycle.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 529c84, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during unknown surgery, the knife did not move and no feedback at the 3rd stroke of the first firing, and the device could not be fired.The device was used on a colon.The cartridge color was blue.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON FLEX 60 COMPACT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key20410280
Report Number3005075853-2024-07616
Device Sequence Number18418312
Product Code GDW
UDI-Device Identifier10705036002710
UDI-Public(01)10705036002710(17)260731(10)529C84
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K081146
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Catalogue NumberSC60A
Device Lot Number529C84
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/25/2024
Initial Report FDA Received Date10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GST60B
Patient SexUnknown
-
-