• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATTUNE IM ROD; ATTUNE INSTRUMENTS : ALIGNMENT DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATTUNE IM ROD; ATTUNE INSTRUMENTS : ALIGNMENT DEVICES Back to Search Results
Catalog Number 254400502
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event is not a valid lot number, the full udi is currently not available.H3, h6: product investigation: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found signs of nicked and scratches on the rod.Additionally, the device presents signs of worn, consistent with the time of service of the attune im rod.The observed condition was identified as an end-of-life indicator, damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune im rod would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (b1011) provided is not a valid finished goods lot#.
 
Event or Problem Description
It was reported that the femoral alignment rod had a scratch in it that the sterile processing department person stated is no longer able to be used in cases.During manufacturer's investigation of the returned device it was identified that device presents signs of worn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE IM ROD
Common Device Name
ATTUNE INSTRUMENTS : ALIGNMENT DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20410686
Report Number1818910-2024-21239
Device Sequence Number13054186
Product Code HWT
UDI-Device Identifier10603295130024
UDI-Public(01)10603295130024
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254400502
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/03/2024
Initial Report FDA Received Date10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-