Model Number 1345-2803
Medical Device Problem Code Wrong Label (4073)
Health Effect - Clinical Codes Joint Dislocation (2374); Osteolysis (2377); Joint Laxity (4526); Metal Related Pathology (4530)
Type of Reportable Event Malfunction
Event or Problem Description
An initial complaint was received on september 12, 2024, indicating that it was discovered intraoperatively that a box labeled as containing a size 28 l ceramic head actually contained a size 32 l ceramic head (correct size directly marked on the device).No patient injury or death was reported.Novosource became aware that biolox ceramic heads, a component in the novosource total hip system had been mislabeled.Sizes 28 l and 32 l had seemingly been reversed.The preliminary investigation indicated that the issue was restricted to 2 lots, each containing a quantity of 10 units.As part of the ongoing complaint investigation and containment actions, the suspect lot and the associated lot (packaging labeled as size 32 l heads) were requested to be returned for evaluation.Upon inspection of the returned units, novosource is aware that the 28 l ceramic heads were indeed also packaged with labeling that indicated size 32 l.Thus, the decision was been made that a firm-initiated recall was appropriate and the appropriate fda division recall coordinator was notified without delay.Finally, information obtained through a health hazard evaluation supported that if the malfunction were to recur, it would likely cause or contribute to serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVOHIP TOTAL HIP SYSTEM
Common Device Name
HIP PROSTHESIS
Manufacturer (Section D)
NOVOSOURCE, LLC
1000 hampton center
suite a
morgantown WV 26505
Manufacturer (Section G)
NOVOSOURCE, LLC
1000 hampton center
suite a
morgantown WV 26505
Manufacturer Contact
jessica singelais
1000 hampton center
suite a
morgantown, WV 26505
3045542167
MDR Report Key20412105
Report Number3014273644-2024-00002
Device Sequence Number7969682
Product Code LPH
UDI-Device Identifier00811382037818
UDI-Public00811382037818
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K132158
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 09/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1345-2803
Device Catalogue Number1345-2803
Device Lot NumberCER0021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2024
Initial Report FDA Received Date10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3014273644/09232024/C/00
Patient Sequence Number1
Patient Age77 YR
Patient SexMale
-
-