BECTON, DICKINSON AND COMPANY (BD) BD PRESET¿ ECLIPSE¿ ARTERIAL BLOOD COLLECTION SYRINGE; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
|
Back to Search Results |
|
| Catalog Number |
364391 |
| Medical Device Problem Code |
Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported when using the bd preset¿ eclipse¿ arterial blood collection syringe, the needles are blunt and short, as well as insufficient blood flow in an unspecified number of syringes.There were no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd had not received samples or photos for investigation.Therefore, (b)(4) retention samples from bd inventory were evaluated by visual examination and functional draw testing and no issues were observed relating to dull needle point, short needle length, or insufficient blood flow as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure modes dull needle point, short needle length, or insufficient blood flow.Bd was not able to identify a root cause for the indicated failure modes.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|