OLYMPUS WINTER & IBE GMBH OP-TELESCOPE, 30°, 4 MM CHANNEL; RIGID NEPHROSCOPE Back to Search Results
Model Number WA33036A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the rigid nephroscope had a broken lens.The issue occurred during reprocessing.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The customer's allegation was not confirmed.The device was not returned for evaluation and could not be evaluated.Based on the results of the investigation, it is possible that the following led to the malfunction: the effect of a large mechanical force due to a shock, fall or collision with other equipment.However, a definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OP-TELESCOPE, 30°, 4 MM CHANNEL
Common Device Name
RIGID NEPHROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20415161
Report Number9610773-2024-02492
Device Sequence Number12826420
Product Code FGA
UDI-Device Identifier04042761051866
UDI-Public04042761051866
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K853486
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA33036A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2024
Supplement Date Received by Manufacturer04/16/2025
Initial Report FDA Received Date10/10/2024
Supplement Report FDA Received Date05/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/16/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-