| Model Number |
53401 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis: it was determined at analysis that the external pulse generator's(epg) rate encoder did not increase or decrease.It was noted that the encoder assembly was out of specification electrical.All found defective parts were replaced and all other identified issues were resolved.The epg then passed all final functional tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
The external pulse generator (epg) was returned for preventive maintenance and subsequently tested out of specification during manuf acturer's analysis.There was no patient involvement.
|
| |
|