IVANTIS INC HYDRUS MICROSTENT; INTRAOCULAR PRESSURE LOWERING IMPLANT Back to Search Results
Catalog Number 100300293
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Intraocular Pressure Increased (1937); UGH (Uveitis Glaucoma Hyphema) Syndrome (4470)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Literature citation: n.Sachdeva et al., early to late explantation of hydrus microstent migs device: a case series: american journal of ophthalmology case reports 2024 jul 20;36:102105, 1-4.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A health care professional published original article with a purpose early to late explantation of hydrus microstent migs device: a case series.The hydrus micros tent was approved by the fda in (b)(6) 2018 for use with cataract surgery to reduce iop in patients with mild to moderate primary open angle glaucoma (poag).Pivotal clinical trials demonstrated its overall safety and efficacy in lowering iop.However, mispositioning of the implant can result in uveitis glaucoma- hyphemia (ugh) syndrome necessitating device explantation.Here we report four such cases and their associated challenges.We also highlight the importance of early recognition of post-operative complications for ease of implant removal.A 71-year-old man with poag on dual anti-platelet therapy (aspirin and clopidogrel) presented with persistent hyphema and uncontrolled iop after the hydrus implant.During his cataract and hydrus implant surgery, significant bleeding was noted from the initial goniotomy site for hydrus insertion, and additional viscoelastic was able to tamponade the bleed intraoperatively.In the post-operative period, the patient developed a large hyphema and nasal clot with an iop of 64 mmhg and was started on oral acetazolamide.Clopidogrel was stopped after consulting the cardiologist, but aspirin 81 mg was continued.The patient continued to have hyphema with uncontrolled iop on maximal medical therapy.B-scan was negative for posterior segment hemorrhage.The patient underwent anterior chamber washout with hydrus explantation and ahmed glaucoma valve implantation 16 days after the first surgery.Intraoperatively, the proximal end of the implant was surrounded by peripheral anterior synechiae and iris tissues.It was pulled free using a sinsky hook, and a hydrus deployer was used to retract the implant into the anterior chamber and out of the eye.On pod 1, iop was 11 mmhg with visual acuity of hand motion and no hyphema, but vitreous hemorrhage was noted.The patient was started on prednisolone acetate four times daily with a gradual taper.By 5 months, the vitreous hemorrhage had cleared with improvement in visual acuity to 20/40, and iop was 15 mmhg on medication.This case highlights that patients with hydrus implants treated with anti-platelet therapy may be at higher risk for persistent hyphema in cases of implant malposition.Conclusion: hydrus micros tents that are mispositioned can result in persistent uveitis-glaucomahyphema syndrome.Explantation between 2 weeks and 11 months successfully resolved post-operative uveitis and hyphemia, but all cases required additional glaucoma-hyphemia syndrome.Early recognition is important since late removal was more challenging due to the implant becoming embedded in the iris.
 
Additional Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.11.A sample was not received at investigation site for evaluation for the report of persistent hyphema, uncontrolled intraocular pressure, microstent explantation, implant was surrounded by peripheral anterior synechiae and iris tissues; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the investigation site and no lot information is available, therefore, the root cause for the customer complaint issue cannot be determined.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDRUS MICROSTENT
Common Device Name
INTRAOCULAR PRESSURE LOWERING IMPLANT
Manufacturer (Section D)
IVANTIS INC
201 technology drive
irvine CA 92618
Manufacturer (Section G)
IVANTIS INC
201 technology drive
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20415701
Report Number3016075957-2024-00047
Device Sequence Number8202781
Product Code OGO
UDI-Device Identifier00867487000134
UDI-Public00867487000134
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P170034
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Literature,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number100300293
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/12/2024
Supplement Date Received by Manufacturer12/09/2024
Initial Report FDA Received Date10/10/2024
Supplement Report FDA Received Date01/02/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-