|
This report has been identified as b.Braun internal report number (b)(4).A follow-up report will be provided after the examination results are available.This report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|
|
This report has been identified as b.Braun internal report number (b)(4).A follow-up report will be provided after the examination results are available.Note: this report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.Corrected information d4 primary unique device identifier (udi) number.Additional informaiton: d9 device returned to manufacturer h4 device manufacturer date.General information: complaint: (b)(4) information to the sample: model: perfusor space article number: 8713030 serial number/batch:(b)(6) software version: n030004 hours of operation: 16408 further information: n/a.Investigation results: history inspection: the device history files were read and analyzed.The device history files from (b)(6) 2024 were analyzed.On (b)(6) 2024 at 12:00am a omnifix 50ml syringe was inserted (filled approx.6,5ml) and the infusion started with a rate of 0,25ml/h and a volume of 10ml.During the infusion a bolus was given, several times.At the next day at 07:51am the syringe was empty.At this time, 6,4ml was infused.A second infusion was started.On (b)(6) 2024 at 08:01am a omnifix 50ml syringe was inserted (filles approx.To 5ml) and the infusion started with a rate of 0,25ml/h and a volume of 10ml.During the infusion a bolus was given, several times.At 17:40pm the syringe was empty and the infusion stopped.At this time, 4,42ml was infused.No other abnormalities were found.Visual inspection: a visual inspection was performed.The cover caps on the screw pillars, and the technician seal (41-01-077) on the lower housing were intact and undamaged.The device shows age related signs of wear and tear but no visible damage was found.Functional inspection: a functional test was performed.The device passes the self-test.A b.Braun omnifix 50 ml syringe was inserted, and the pump identified the syringe, and it could be selected from the menu.It was possible to put the pump in operation.Individual inspection: a delivery accuracy measurement according to iec 60601-2-24 was arranged.Here a nominal flow rate of 5 ml/h was chosen.The assessed mean deviation "a" of the second operating hour was measured and resulted in a value of +0,96%.Accuracy of set delivery rate should be ± 2 % according to iec/en 60601-2-24.Disassembling: during the investigation no faults could be detected, to investigate the inside of the device, only the upper housing was removed.No damage or soiling could be found.Judgment: the complaint could not be confirmed.Summing up all tests, the perfusor space operated within our specification.No product deviation.
|