| Model Number |
320574 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Spouse of consumer reported needle clog during injection.He stated that there is no insulin flow.Consumer also reported non patient end needle bent.Issues involve two different lots.Consumer does not prime, consumer does not re-use.Lot #: 3115886, 3179396.Catalog #: 320574.Date of event: unknown.Samples available: sending mail kit.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to: h6 (component code, investigation findings, and investigation conclusions) investigation summary: samples were received and an investigation was performed.This is the 2nd complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformance's were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue as bent/broken cannula and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
| |
|