• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND PEN NDL 32G 4MM PRO 90CT US; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND PEN NDL 32G 4MM PRO 90CT US; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number 320574
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Spouse of consumer reported needle clog during injection.He stated that there is no insulin flow.Consumer also reported non patient end needle bent.Issues involve two different lots.Consumer does not prime, consumer does not re-use.Lot #: 3115886, 3179396.Catalog #: 320574.Date of event: unknown.Samples available: sending mail kit.
 
Additional Manufacturer Narrative
Correction to: h6 (component code, investigation findings, and investigation conclusions) investigation summary: samples were received and an investigation was performed.This is the 2nd complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformance's were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue as bent/broken cannula and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NDL 32G 4MM PRO 90CT US
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key20417787
Report Number3023359743-2024-00546
Device Sequence Number8401194
Product Code FMI
UDI-Device Identifier00382903205745
UDI-Public00382903205745
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K212015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number320574
Device Lot Number3115886
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/10/2024
Supplement Report FDA Received Date12/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-