|
Additional information: g3, h3, h6.Based on the information provided, which may include the device (if available and returned), pictures, videos, event descriptions, and any additional information from the field, arthrex was able to conclude the most likely cause.The most likely cause(s) of the reported failure can be user error, such as misaligned insertion, prying/leveraging, and/or applying excessive force on the device during use.Direction for use.Dfu-0088-2 at revision 0.Monopolar ablation probe.Section c.Contraindications.2.Use during an arthroscopic procedure without a conductive irrigation solution.Section d.Adverse effects.3.Premature probe tip wear may be caused by vigorous use against bony surfaces.Section f.Warnings 4.Do not use non-conductive irrigation fluids (sterile water, air, gas, glycine, purisole, sorbitol-mannitol, etc.).6.Do not insert or withdraw the probe tip from the surgical site while the device is active.9.Do not make probe contact with an arthroscope during activation as it may damage the probe and/or scope.11.Excessive wear of electrodes may result from vigorous use against bony tissue.The complaint allegation was confirmed based on the customer's attached picture (251785 b.Png) which displays part of the ablator' s tip broken off.
|