It was reported, the deflectable videoscope exhibited peeling of the light guide lens adhesive.The issue is unknown as to when it occurred.There were no reports of delay, patient harm, injury, or death.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer¿s final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Based on the results of the investigation, the definitive root cause of the reported issue could not be determined, however, it is likely physical stress from bumping the distal end of the scope or dropping the distal end, or chemical stress from the chemical solution used led to the malfunction.The event can be detected/prevented by following the instructions for use which state: operation manual for ltf-190-10-3d important information ¿ please read before use dangers, warnings, and cautions: do not strike, hit, or drop the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Reprocessing manual for ltf-190-10-3d 3.1 compatibility summary.Methods listed as ¿compatible¿ in table 3.1 and table 3.2 are compatible for routine use only when used according to manufacturer¿s instructions.Repeated use and reprocessing of endoscopes and accessories lead to gradual wear and tear.Furthermore, reprocessing methods that employ higher temperatures and more caustic/corrosive materials may led to faster deterioration.In general, sterilization processes are harsher on equipment than disinfection processes.Before each patient procedure, inspect the endoscope and accessories for damage, according to the instructions described in this manual and its companion ¿operation manual¿.3.1 compatibility summary: ¦list of compatible methods validated in terms of biological efficacy and material durability sterilization by sterrad®100s/nx®/100nx® system may deteriorate the adhesive of the insertion section.Depending on the circumstances, replacement of the insertion section may be required.Before use, confirm that the endoscope is free from any damage or other irregularities.For further details, contact olympus.Olympus will continue to monitor field performance for this device.
|