• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIBIA TRAPTRAY CEM SZ 2F/2T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIBIA TRAPTRAY CEM SZ 2F/2T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-41-2020
Medical Device Problem Codes Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Scar Tissue (2060); Inadequate Osseointegration (2646); Metal Related Pathology (4530)
Date of Event 08/26/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitants: (b)(6) 02-012-44-2009 - logic tibia implant psc insert, sz 2, 9mm (b (6) 02-020-11-0320 - truliant ps cem fem ps cem right sz 2 (b)(6) 200-02-35 - three peg patella 35mm.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
 
Event or Problem Description
Legal case ¿ usa it was reported that approximately 67 months after a right total knee replacement procedure, the patient underwent a revision procedure to address massive poly wear.Revision surgery operative notes were provided.Indication: aseptic loosening in the setting of massive poly wear from recalled implant.Procedure description: it was noted that there was hypertrophic and metallic appearance of the synovium.An examination of the polyethylene liner revealed catastrophic failure of the medial compartment leading to metal on metal articulation.Retractors were placed and the femoral component was removed with minimal bone loss.Next the tibia was presented anteriorly and a thin curved osteotome was used to remove the tibial component.This again was performed with minimal difficulty and minimal bone loss.There was a significant cavitary lesion in the patella that was debrided with a rongeur and curette.The patient was transferred to pacu in stable condition.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: a, b, e.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA TRAPTRAY CEM SZ 2F/2T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville 32653
Manufacturer Contact
geoff gannon
gainesville, FL 32653
MDR Report Key20422969
Report Number1038671-2024-03933
Device Sequence Number8735607
Product Code JWH
UDI-Device Identifier10885862003119
UDI-Public10885862003119
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/21/2019
Device Catalogue Number02-012-41-2020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2024
Supplement Date Received by Manufacturer05/13/2025
Initial Report FDA Received Date10/10/2024
Supplement Report FDA Received Date05/14/2025
Date Device Manufactured07/22/2011
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-