• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN MAR +4 NEUT 36IDX58OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN MAR +4 NEUT 36IDX58OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 121936458
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 10/03/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was revised on (b)(6) 2024 due to painful right tha.Surgeon revised head and liner and removed screws.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 NEUT 36IDX58OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20423481
Report Number1818910-2024-21349
Device Sequence Number18326806
Product Code LPH
UDI-Device Identifier10603295014737
UDI-Public(01)10603295014737
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K010171
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number121936458
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/03/2024
Supplement Date Received by Manufacturer10/25/2024
Initial Report FDA Received Date10/10/2024
Supplement Report FDA Received Date10/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexUnknown
-
-