MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SMPX645 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: as of now there is no information regarding the explant or reoperation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 30-year-old female patient underwent primary breast augmentation with 645cc mentor memorygel xtra breast implant and experienced capsular contracture; baker grade iii/iv on her left side postoperatively; diagnosed in the office.The patient has consulted with the healthcare professional.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
|
| |
|
Additional Manufacturer Narrative
|
|
On october 24, 2024, mentor became aware that the date of surgery is (b)(6) 2024.Hence, following fields have been updated on this form: - is required intervention has been selected under section b; field b2.- fields d6b for explantation date is (b)(6) 2024.- field h6 health effect - impact code ¿device explantation¿ has been added.- field h6 type of investigation has been updated to "device not returned".Reason for device explant and/or reoperation: left capsular contracture; baker grade iii/iv.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On december 5, 2024, the mentor failure analysis lab received the device for evaluation.On december 18, 2024, device evaluation was completed as follows: mentor conducted a visual inspection of the returned device.During visual evaluation no apparent damage or visual anomalies were observed on the smooth mod + prof xtra 645cc returned device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On november 11, 2024, mentor became aware that the patient underwent left side implant exchange surgery with exactly the same device on (b)(6) 2024.Manufacturer¿s reference number: (b)(4).
|
| |
|