MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. SAFEDAY¿; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. SAFEDAY¿; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 352905
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(6).One used sample in open packaging and 126 unused samples in original packaging were provided for evaluation.All samples were visually evaluated, and no defects were observed.As per aql 80 samples were leak tested and leakage was detected at the bonded joint of the female luer lock connector and safeday valve.Further investigation revealed insufficient solvent at the bonded joint.Leakage was not identified on the other 79 unused samples tested.Based on the evaluation results, the reported defect was confirmed.Incidents of this nature are attributed to operator oversight during the manual solvent application process of the product.The operator applied insufficient solvent on the tubing during the bonding process.Although our training procedures ensure that our employees are properly trained in their areas of responsibility, an oversight on the operator's part can be attributed to an incident of this nature.Retained units were evaluated and passed the internal testing.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.Note: the original initial mdr for this case was inadvertently submitted via the "test" account of the fda electronic submission gateway (esg) webtrader.As the original submission was not submitted through the "production" account of the fda esg webtrader, a new initial mdr submission is required per zoom call with (b)(6) of the fda esg help desk.
 
Event or Problem Description
As reported, event 1: recently we have had at least three instances where our j-loops for iv's, safeday extension set ref 352905 lot 0061938553 have been leaking from the connection to the clave.These extension sets are sometimes used with patients who are receiving chemotherapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDAY¿
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd
allentown PA 18109
Manufacturer (Section G)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km 22 autopista las americas
santo domingo,
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key20427291
Report Number2523676-2024-00818
Device Sequence Number15350801
Product Code FPA
UDI-Device Identifier04046955678129
UDI-Public(01)04046955678129
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K153293
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number352905
Device Lot Number0061938553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2024
Initial Report FDA Received Date10/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1