|
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus for inspection, and the reportable malfunction was confirmed.Based on the results of the investigation, it is likely the following led to the malfunction: 1.The grasping section was closed without grasping anything while the output activation in seal & cut mode was activated, (this includes after tissue resection) causing the tissue pad to wear out and peel off partially.2.Due to wear of the tissue pad, the non-insulated area of the grasping section came into contact with the probe.3.The output was activated while the non-insulated area of the grasping section was contacting the probe causing a probe damage error.As a result, a contact mark developed.However, a definitive root cause could not be identified.The event can be detected/prevented by following the instructions for use which state: the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, other instruments, or forceps, and others.Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/split/protruding/partial separating of the tissue pad.In turn, the probe tip may break before displaying an error window or generating an alarm tone.During output, do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments.Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities, which may lead to burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.Do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.Olympus will continue to monitor field performance for this device.
|