| Model Number |
DISPOSABLE KIT, KNEE FIBERTAKS |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 10/03/2024, it was reported by a sales representative via (b)(4) that an ar-3710 disposables kit drill guide was producing debris.This was discovered during a case with no patient harm.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or prying/leveraging the device during use.
|
| |
|