AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a2, a3, a4, a5 and a6: unknown/ not provided.Section d6b: if explanted, give date: not applicable, as there is no indication that the lens was explanted section e1: reporter: middle name: unknown/ not provided section e1: reporter: last name: unknown/ not provided section e1: reporter: email address: unknown/ not provided section e1: reporter: address: address - line 2: unknown/ not provided section e1: reporter: address: city: unknown/ not provided section e1: reporter: address: state, province, or territory: unknown/ not provided section e1: reporter: address: post office or zip code: unknown/ not provided section e1: reporter: telephone number:(b)(6).Device evaluation: at the time of the investigation, device was not available for evaluation, therefore no testing could be performed.The reported event cannot be confirmed.Manufacturing record review: the manufacturing records for the device were reviewed.The product was manufactured and released according to specification.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the haptic of the monofocal non preloaded intraocular lens (iol) was broken.Issue was observed after implantation of the iol.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key20428953
Report Number3012236936-2024-000275
Device Sequence Number15554206
Product Code HQL
UDI-Device Identifier05050474531673
UDI-Public(01)05050474531673(17)280604
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZCB00
Device Catalogue NumberZCB0000180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2024
Initial Report FDA Received Date10/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-