MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LATITUDE PROGRAMMING SYSTEM; EXTERNAL PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LATITUDE PROGRAMMING SYSTEM; EXTERNAL PROGRAMMER Back to Search Results
Model Number 3300
Medical Device Problem Code Interrogation Problem (4017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient with a implantable cardioverter defibrillator (icd) was to undergo a magnetic resonance imaging (mri) so it was to be programmed into mri protection mode.The pacemaker was not able to be interrogate with the programmer, with it displaying a wand out of range message.Technical services (ts) performed troubleshooting, with the issue not resolved so the caller was advised to contact their local field representative.This device remains in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that the patient with a implantable cardioverter defibrillator (icd) was to undergo a magnetic resonance imaging (mri) so it was to be programmed into mri protection mode.The pacemaker was not able to be interrogate with the programmer, with it displaying a wand out of range message.Technical services (ts) performed troubleshooting, with the issue not resolved so the caller was advised to contact their local field representative.This device remains in service.No adverse patient effects were reported.Additional information from the field indicates the issue was resolved, a new programmer and a new wand was used and it was able to interrogate the icd.The issues were attributed to damage in the wand.This device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE PROGRAMMING SYSTEM
Common Device Name
EXTERNAL PROGRAMMER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
PLEXUS ELECTRONICA S DE RL DE CV
paseo del norte no
4640 guadalajara
zapopan, jal 45010
MX   45010
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key20428957
Report Number2124215-2024-63669
Device Sequence Number12862566
Product Code OSR
UDI-Device Identifier00802526610110
UDI-Public00802526610110
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P910077/S159
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3300
Device Catalogue Number3300
Device Lot Number010960
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2024
Supplement Date Received by Manufacturer10/11/2024
Initial Report FDA Received Date10/11/2024
Supplement Report FDA Received Date10/14/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-