MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3542640 |
| Medical Device Problem Codes |
Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
10/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast augmenation surgery with implantation of a 275cc mentor siltex round moderate profile saline breast implant prosthesis.Post-operatively, the patient was diagnosed with bilateral baker grade ii-iii capsular contracture of the breasts and suspected implant deflations, bilaterally, by a medical professional.As a result, the patient underwent bilateral removal and replacement with 375cc mentor memorygel boost breast implants on (b)(6) 2024.This report is for the right breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.D4: udi: product made prior to udi compliance date.Udi not required.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: capsular contracture, suspected deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2024, the mentor analysis lab received the suspect medical device for evaluation.Paperwork received with the device indicated only the left implant as possibly deflated.This report is for the right breast prosthesis.As this report is for the right side, deflation is no longer applicable to this file.On (b)(6) 2024, mentor was informed at the patient only experienced a left deflated implant and right breast capsular contracture.On (b)(6) 2024, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection and leak testing of the device.During visual evaluation no apparent damage or visual anomalies were observed on the breast implant device.Leak testing was performed, in accordance with mentor procedures, and no leak sites were detected during the analysis.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.Manufacturer¿s reference number: (b)(4).
|
| |
|