It was reported that a progav shunt system (#fv415t) was implanted during a procedure performed on (b)(6) 2022.According to the complainant, the shunt system showed an under-drainage.The patient underwent a revision procedure performed on (b)(6) 2024.The complainant device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 9 years, 6 months.Weight: 20.6 kgs.Height: 115 cm.Gender: female.
|
|
Visual inspection: during the investigation, a deformation of the outer housing of the progav was determined, the shuntassistant was without abnormalities.Permeability test: a permeability test has shown that all components are permeable.During the permeability test bloody liquid were flushed out.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves do not operate within the permissible tolerance in their respective relevant positions.An accelerated outflow of both valves could be determined.Adjustment test: the progav was tested and is not adjustable.Raking force and brake function test: the brake functionality test has shown that the brake function operates as expected; however, the breaking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, deposits were found inside.Results: based on our investigation results, we can determine an accelerated outflow at both valves in their respective relevant positions as well as non-adjustability at the progav.He deposits visible in the valves and the deformation of the housing surface on the progav have led to the functional impairments.Deposits caused by natural substances in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of proteins can impair the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|