• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT SZ6/7 TAPERED REAMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT SZ6/7 TAPERED REAMER Back to Search Results
Catalog Number 257002300
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported on (b)(6) 2024 that the hospital staff dropped the bipolar retaining clip.The summit reamers have become dull.This did not delay surgery.
 
Additional Manufacturer Narrative
Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that the hospital staff dropped the bipolar retaining clip on the floor and threw it in the trash.Also, the summit reamers have become dull.This did not delay surgery.No other information is available.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the device found blunting/rounding of cutting edges.It is not unreasonable that the condition identified in visual analysis would contribute to a dull condition.Therefore, we are able to confirm dullness with damage observed, as this kind of evidence indicates repeated use of the device.The lifecycle requirements of the device are event related and depend on the use in clinical practice, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of summit sz6/7 tapered reamer would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT SZ6/7 TAPERED REAMER
Common Device Name
REAMER
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key20429431
Report Number1818910-2024-21429
Device Sequence Number8735808
Product Code HTO
UDI-Device Identifier10603295143024
UDI-Public(01)10603295143024(10)SO2013047
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number257002300
Device Lot NumberSO2013047
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/04/2024
Supplement Date Received by Manufacturer11/06/2024
11/13/2024
Initial Report FDA Received Date10/11/2024
Supplement Report FDA Received Date11/07/2024
11/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
-
-