• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 MICROEZ MICROINTRODUCER KIT 5F (7CM, NITINOL); INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 MICROEZ MICROINTRODUCER KIT 5F (7CM, NITINOL); INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by customer that nine picc kit introducer had issues with difficult insertion.No other information was provided.09/12/2024: nine 5fr microez microintroducers were returned for evaluation, it was found during an investigation there five of the microintroducers had deformation of the sheath.This report addresses the second device.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The complaint of a difficult microintroducer insertion was confirmed and determined to be use related.The products returned for evaluation were nine 5fr microez microintroducers.The returned units were evaluated and deformation of the sheath was observed on all nine devices.The following observations were noted during the sample evaluation: usage residue was seen on all returned units.Damaged at the edges of the sheath was present in units 01, 04, 05, 08, and 09.Buckling of the sheath, which is indicative of compression failure, was observed at the location of bends in the sheath in units 02, 03, 06, and 07.The shape, location, and extent of the damage observed on the returned units suggested that the microintroducer sheaths were most likely damaged due to excessive insertion force, an inadequate skin nick, and/or a steep insertion angle.This complaint will be recorded for future trending and monitoring purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROEZ MICROINTRODUCER KIT 5F (7CM, NITINOL)
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key20430012
Report Number3006260740-2024-05869
Device Sequence Number7860778
Product Code DYB
UDI-Device Identifier00801741038440
UDI-Public(01)00801741038440(11)230628(17)250430(10)REHS0315
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K915167
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model NumberN/A
Device Catalogue Number0928950
Device Lot NumberREHS0315
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2024
Initial Report FDA Received Date10/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-