| Catalog Number |
AB2000 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Root cause of the reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|
| |
|
Event or Problem Description
|
|
A male patient underwent aquablation therapy for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that the aquabeam robotic system generated an "e22 - motorpack error" when connecting the aquabeam handpiece to the aquabeam motorpack.The e22 error was cleared.During nozzle registration the aquabeam robotic system generated another "e22 - motorpack error".The error was again cleared.During the first treatment pass, the aquabeam robotic system generated multiple "e22 - motorpack error", "e23 - motorpack error", and "e31 - motorpack error".After troubleshooting, the first treatment pass was able to be completed.During the second treatment pass, the aquabeam robotic system generated multiple "e22 - motorpack error", "e23 - motorpack error", and "e31 - motorpack error".As a result, the treating surgeon decided to use a bipolar loop to finish the treatment.There were no adverse health consequences to the patient due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The aquabeam handpiece was not returned for investigation of the reported event as it was disposed of at the hospital.A replacement aquabeam handpiece was provided to the hospital as part of warranty replacement.The treatment log files were reviewed and found that the system triggered an e22 error along with an e23 error, confirming the reported event.The root cause of the reported event was unable to be established as the device was not returned for investigation.A review of the device history record (dhr) for aquabeam robotic system/serial number (b)(6) and the aquabeam handpiece / lot number 23c03489 was performed, which confirmed that there was one (1) non-conformance issued to this lot during the manafacturing process that could potentially be related to the reported failure.The lot was segregated and affected units were reworked and re-inspected as part of our handpiece final inspection process.Upon re-inspection, the lot met all required specifications and was deemed acceptable to be released for distribution.The current user manual um0101-00 rev.F, aquabeam robotic system user manual was reviewed and states the following: table 5: system detected errors and faults e22 - motorpack error release foot pedal and click x.1) if error persists, reconnect handpiece to motorpack.2) if error continues, replace handpiece.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|
| |
|
Search Alerts/Recalls
|